This research introduces a novel control framework for robotic ultrasound that leverages image-based contact perception to eliminate reliance on external force sensors. By dynamically updating the interaction frame based on visual servoing, the system significantly reduces compressive forces while maintaining high tracking accuracy during probe manipulation. The work represents a tangible advancement in safe human-robot interaction within medical imaging applications.
Contact-Aware Impedance Controller for Robot-Assisted Ultrasound Imaging
arXiv:2609.34400v1 Announce Type: new Abstract: Safe robot-assisted ultrasound imaging requires a reliable controller able to detect and localize probe--tissue interaction. In this paper, we present a B-mode ultrasound image-based contact perception method and a contact-aware impedance controller for robotic ultrasound imaging. The proposed method detects acoustic contact independently of force measurements, enabling contact-conditioned force/torque taring to reduce residual wrench bias. During contact, the method continuously estimates the effective contact location along the curved probe surface and uses it to update the controller interaction frame, enabling visual servoing of the physical probe--tissue contact point during imaging. Experiments on an agar phantom demonstrated a contact-localization RMSE of $\mathbf{1.46 \pm 0.14}$~mm over probe roll angles from $\mathbf{-15^\circ}$ to $\mathbf{15^\circ}$. During static rolling, the proposed controller maintained task-space tracking accuracy comparable to a conventional fixed-frame impedance controller while reducing the maximum compressive interaction force from $\mathbf{31.56}$~N to $\mathbf{20.09}$~N, corresponding to a $\mathbf{36.3\%}$ reduction. These results demonstrate the potential of ultrasound images as direct contact feedback for safe and accurate robot-assisted ultrasound imaging.
Abstract:Safe robot-assisted ultrasound imaging requires a reliable controller able to detect and localize probe--tissue interaction. In this paper, we present a B-mode ultrasound image-based contact perception method and a contact-aware impedance controller for robotic ultrasound imaging. The proposed method detects acoustic contact independently of force measurements, enabling contact-conditioned force/torque taring to reduce residual wrench bias. During contact, the method continuously estimates the effective contact location along the curved probe surface and uses it to update the controller interaction frame, enabling visual servoing of the physical probe--tissue contact point during imaging. Experiments on an agar phantom demonstrated a contact-localization RMSE of $\mathbf{1.46 \pm 0.14}$~mm over probe roll angles from $\mathbf{-15^\circ}$ to $\mathbf{15^\circ}$. During static rolling, the proposed controller maintained task-space tracking accuracy comparable to a conventional fixed-frame impedance controller while reducing the maximum compressive interaction force from $\mathbf{31.56}$~N to $\mathbf{20.09}$~N, corresponding to a $\mathbf{36.3\%}$ reduction. These results demonstrate the potential of ultrasound images as direct contact feedback for safe and accurate robot-assisted ultrasound imaging.
Robotics (cs.RO)
arXiv:2609.34400 [cs.RO]
(or arXiv:2609.34400v1 [cs.RO] for this version)
https://doi.org/10.48550/arXiv.2609.34400 Focus to learn more arXiv-issued DOI via DataCite (pending registration)
arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.
Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.
Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.
Text integral sub licența deschisă a publicației · 256 cuvinte
Subiecte conexe
Contact-Adaptive Robotic Ultrasound Probe Control for Tissue Exploration and Continuous Task-Relevant Visualization Using Robot-Free Image-Motion Demonstration
arXiv · 30 sept. 2026
Design and Evaluation of a Compliant Quasi Direct Drive End-effector for Safe Robotic Ultrasound Imaging
arXiv · 26 aug. 2026

The Machines that Make the Machines
NVIDIA · 7 oct. 2026
Transferable Tool-Tissue Contact Detection from Stereo Depth in Robot-Assisted Surgery
arXiv · 20 aug. 2026