Skip to content
Off-topic for aiqbench

Off-topic for aiqbench

This story was gathered from a source we follow but falls outside AI and quantum coverage, so there is no aiqbench take for it.

Physically Constrained Kinematic Calibration for High-Precision Robotic Needle Guidance

arXiv2 min read

arXiv:2610.41429v1 Announce Type: new Abstract: Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.

Abstract:Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.

Robotics (cs.RO); Systems and Control (eess.SY); Optimization and Control (math.OC)

I.2.9; J.3; G.1.6

arXiv:2610.41429 [cs.RO]

(or arXiv:2610.41429v1 [cs.RO] for this version)

https://doi.org/10.48550/arXiv.2610.41429 Focus to learn more arXiv-issued DOI via DataCite (pending registration)

arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.

Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.

Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.

Full text under the publisher's open licence · 292 words